Accuracy, High Performance And Productivity At An Affordable Price

With the DataStar product family you can have it all: speed, high accuracy, full features, ease of use and outstanding value. The DataStar series are built on a solid granite base, optimized for stable, long-term performance for the most demanding applications. This compact bench-top system features ROI Optics, LED bottom light, fiber optic transmitted ring light and on-axis light.

The DataStar is powered by the AutoCheck® and easy to use Basic-X® Measurement Software. With an extensive array of measurement tools, including the AutoFocus™ Motorized Zoom, intelligent edge detection, user friendly icons and a point and click interface, Basic-X makes it simple for everyone to get accurate, repeatable measurements. AutoCheck is compatible with most Windows-based operating systems.

X-Y AXIS English Units Metric Units
X-Y Axis Travel 200 9" X 6" 225 x 150 mm
X-Y Axis Travel 300 12" X 6" 300 x 150 mm
X-Y Axis Linear Accuracy* E2=140 + 10L (in) µinch E2=3.5 + L (mm)/ 100 µm
Scale Resolution .00002" 0.5 µm
Z-AXIS
Z Axis Travel 6" 150 mm
Z Axis Accuracy* E1=160 + 10L (in) µinch E1=4 + L (mm)/ 100) µm
Scale Resolution .00002" .05 µm
OPTICAL PERFORMANCE WITH 1X LENS 
Working Distance 2.125" 54 mm
Magnification on 17" Monitor 35X - 180X 35X -180X
Field of View 0.37" - 0.07" 9.4 - 1.8 mm
Zoom Ratio 5X 5X
OPTICAL PERFORMANCE WITH 2X LENS
Working Distance 1.125" 30 mm
Magnification on 17" Monitor 55X - 290X 55X - 290X
Field of View 0.22" - 0.041" 5.6 - 1.04 mm
Zoom Ratio 5X 5X
PHYSICAL
Working Distance 1.125" 30 mm
ENVIRONMENTAL
Temperature, safe operating 5ºC - 40ºC, 41ºF - 104ºF
Temperature to meet specs 20ºC +/- 1ºC, 68ºF +/- 2ºF

*Applies to thermally stable system in rated environment, maximum zoom lens setting, and evenly distributed 5 kg load.

Depending on load distribution, accuracy at maximum rated load may be less than standard accuracy.

XY axis artifact: QVI 25 intersection grid reticle at standard measuring plane.

The standard measuring plane is defined as a plane that is 1" off of the work table. Z axis artifact: QVI step gage.

E1=Z measurement accuracy combines Autofocus and Z axis positioning capabilities.