Economical, Durable Choice For Demanding Part Inspection And Measurement.

The Starlite was designed to be installed on the user's benchtop, and to meet demanding optical inspection and measurement needs accommodating parts up to 13". Starlite comes with three Z options: 3", 6" or our popular 12". The Starlite is powered by the Q-Check® Digital Readout, the Metronics QC200 digital readout, or by the optional AutoCheck® with Basic-X® software.

The Q-Check is a powerful, microprocessor-based measurement system that features XYZ computational capability, measurement memory and print options.

Basic-X was designed specifically for systems with motorized stages and features the AutoFocus™ Motorized Zoom, intelligent edge detection, user friendly icons and a point and click interface.

X-Y AXIS English Units Metric Units
X-Axis Travel 13" 330 mm
Y-Axis Travel 11" 280 mm
X-Y Accuracy* ± .0003"/ inch ± .003 mm / cm
Speed of motorized motion 2.5" / sec 65 mm/ sec
Encoder resolution 0.00004" 1 µm
Z-AXIS
Z Axis Accuracy 6"* ± .0003"/ inch ± .003 mm / cm
Z Axis Accuracy 12"* ± .0005"/ inch ± .005 mm / cm
OPTICAL PERFORMANCE WITH Q-CHECK
Working distance 3.62" 90 mm
Magnification on 13" monitor 35X to 210X 35X to 210X
Field of view 0.322" to 0.053" 8 to 1.35 mm
Zoom ratio 6X 6X
OPTICAL PERFORMANCE WITH OPTIONAL AUTOCHECK
Working distance 3.4" 90 mm
Magnification on 17" monitor 41X to 242X 41X to 242X
Field of view 0.273" to 0.046" 6.94 to 1.176 mm
Zoom ratio 6X 6X
PHYSICAL
External dimensions (H x W x D) 28.8" x 24.7" x 27.3" 730 x 625 x700 mm
Metrology unit weight 150 lbs 70 kg
Maximum workpiece weight 110 lbs 50 kg
Base area (W x D) 524.7" x 24.2" 630 x 615 mm
Bottom illumination fluorescent flat panel
Top illumination fiberoptic ring light
ENVIRONMENTAL
Temperature, safe operating 5ºC - 40ºC, 41ºF - 104ºF
Temperature to meet specs 20ºC +/- 1ºC, 68ºF +/- 2ºF

*Applies to thermally stable system in rated environment, maximum zoom lens setting, and evenly distributed 5 kg load.

Depending on load distribution, accuracy at maximum rated load may be less than standard accuracy.

XY axis artifact: QVI 25 intersection grid reticle at standard measuring plane. The standard measuring plane is defined as a plane that is 1" off of the work table.

Z axis artifact: QVI step gage.